DEFIB

  • Operating Mode : Manual
  • Waveform : Biphasic
  • Energy : 2 to 200 joules in 12 Steps [2, 3, 5, 7, 10, 20, 30, 50, 70, 100, 150, 200 Joules]
  • Adult and Paediatric paddles support
  • Charging Time: less than 8sec
  • Synchronous: Energy delivery begins within 60ms of the QRS Peak
  • Paddle Assembly: Energy paddle assembly with coil retractable high voltage cable
  • Adults & Paediatric paddle integrated to same handle
  • Check mode: Manual test mode

ECG

  • Frequency response : 0.5Hz to 35Hz
  • Patient Isolation : Optical isolation
  • Lead Selection : Lead I, II, III
  • Gain Setting : 0.5, 1 and 2
  • CMRR: 90 dB
  • Sweep Speed: 12.5, 25.0, 50.0mm/sec
  • Waveform 1 : ECG
  • Waveform 2: 4 sec delayed [Cascaded]
  • Digital Filter : Digital Signal processing for good quality of ECG
  • Freeze Mode : To freeze up to 4 seconds ECG

SpO2 (Optional)

  • Range : 0 ~ 100%
  • Accuracy : 70 ~ 100%, ±2%, and 50 ~ 69% ±3%
  • Pulse rate : displayed when PR is deselected
  • Pulse Beep : Step less volume control
  • PR Range: 10 ~ 250 bpm
  • Pitch Variation with change in oxygen saturation

EtC02 [Optional]

  • Parameters : CO2
  • Measurement Unit : % or mm Hg
  • Breathe rate : 3 ~ 150 bpm
  • Range : 0 to 13%
  • Accuracy : 2mmHg < 5.0%CO2[at ATPS]
  • Resolution : 1 mm Hg

NIBP (Optional)

  • Method : Oscillometric
  • NIBP Interval : User Selectable — 3 min to 120 min
  • Modes : Manual, Auto
  • Cuff Deflation : Automatic
  • Safety Features : Automatic deflation if cuff pressure exceeds 280 mmHg
  • Range : 10 ~ 250 mmHg
  • NIBP Status : “Ready” —When not taking reading “Measuring”—When taking reading “Error” — if reading could not be taken due to some error while measuring
  • Patient Type : Adult, Paediatric, Neonate

AED Mode (Optional)

  • Waveform: Bi-Phasic
  • AEC shock series: Energy level from 150 to 200J Charge Command Automatic after identifying arhythmia
  • Shock Command : front panel button,shock
  • Charging Time : less than 8 sec
  • Controls : Analyze, Pause
  • Defibrillation Electrode: Malfunction, Self Adhesive, Adult electrodes
  • Voice and Text Prompts: Voice and Text prompts to guide user

Pacemaker [Optional]

  • Waveform : Monophasic
  • Modes : Demand or Non Demand
  • Amplitude : From 10mA to 15OmA [resolution of 10mA]
  • Pulse Width : 40ms [±1ms]
  • Frequency : from 20ppm to 250ppm [increments of 1ppm]
  • Refectory period: 340ms [from 30 to 80ppm];
    240ms [from 90to 180ppm]
  • Maximum Output Voltage: 150V
  • Impedance Range: 200Ω to 1kΩ

Product Description

  • Display : 7″ Colour TFT
  • Input : 150 ~ 276VAC, 50 Hz
  • Output : 17V ~ 4A
  • Battery : 11.1V, 4Ah Lithium Battery, Rechargeable
  • LED Indicator : Low battery and battery Charging
  • Power Consumption : 30 Watts [Max] in ECG mode
  • Operating Temp : —5°C to + 50°C
  • Humidity : up to 95% non condensing